What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
1 more codes are in the record below.
SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198
- Lot or code
- UDI/DI 4050621100401, 19-0990 UDI/DI 4050621100012, 19-1062 UDI/DI 4050621100432, 19-1152 UDI/DI 4050621100449, 19-1119 UDI/DI 4050621100470, 19-1155 UDI/DI 4050621100487, 19-1156
- FDA recall number
- Z-0329-2025
- Quantity
- 77 units
Distribution (FDA): US Nationwide distribution in the states of NC.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- WISSNER-BOSSERHOFF GMBH
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.