Skip to content
FDAMedical devicesClass IOngoing

FDA recall 95486: Philips Respironics, Inc.: Trilogy Evo O2. Model Numbers: FP2100X10 and 6 more

FDA recall 95486 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
Text from FDA, unedited.

FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

58 more codes are in the record below.

  • Trilogy Evo O2. Model Numbers: FP2100X10, DS2100X11B, IN2100X15B, DE2100X13B, IA2100X15B, EU2100X15B, LA2100X15B, EE2100X15B, FX2100X15B, ES2100X15B, ND2100X15B, FR2100X14B, IT2100X21B, JP2100X16B, IN2100X19, EU2100X19. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.

    Lot or code
    Model Number (UDI-DI Number): DE2100X13B (606959054103); DS2100X11B (606959051997); EE2100X15B (606959055698); ES2100X15B (606959054097); EU2100X15B (606959055674); EU2100X19 (606959055681); FP2100X10 (606959062085); FR2100X14B (606959055643); FX2100X15B (606959061019); IA2100X15B (606959055636); IN2100X15B (606959054059); IN2100X19 (606959055575); IT2100X21B (606959055438); JP2100X16B (606959055582); LA2100X15B (606959055599); ND2100X15B (606959055650). All Serial Numbers.
    FDA recall number
    Z-0290-2025
    Quantity
    24,249 units
  • Trilogy Evo. Model Numbers: DS2110X11B, UDS2110X11B, RDS2110X11B, CA2110X12B, JP2110X16B, IN2110X15B, DE2110X13B, AU2110X15B, BL2110X15B, EU2110X15B, BR2110X18B, LA2110X15B, EE2110X15B, FR2110X14B, GB2110X15B, ND2110X15B, ES2110X15B, IA2110X15B, IT2110X21B, KR2110X15B, TR2110X15B, EU2110X19, FP2110X10, PP2110X10, RBRBR2110X18B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.

    Lot or code
    Model Number (UDI-DI Number): AU2110X15B (606959051980); BL2110X15B (606959055520); BR2110X18B (606959055476); CA2110X12B (606959052307); DE2110X13B (606959054110); DS2110X11B (606959051942); EE2110X15B (606959055544); ES2110X15B (606959054080); EU2110X15B (606959054981); EU2110X19 (606959055537); FP2110X10 (606959062092); FR2110X14B (606959055506); GB2110X15B (606959054127); IA2110X15B (606959055490); IN2110X15B (606959051959); IT2110X21B (606959055421); JP2110X16B (606959051973); KR2110X15B (606959055483); LA2110X15B (606959055452); ND2110X15B (606959055513); PP2110X10 (606959054035); RBRBR2110X18B (606959055476); RDS2110X11B (606959060517); TR2110X15B (606959055551); UDS2110X11B (606959060500). All Serial Numbers.
    FDA recall number
    Z-0291-2025
    Quantity
    82,800 units
  • Trilogy Evo Universal Ventilator. Model Number DS2000X11B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.

    Lot or code
    Model Number DS2000X11B, UDI-DI Number: 606959052000. All Serial Numbers.
    FDA recall number
    Z-0292-2025
    Quantity
    9,999 units
  • Trilogy EV300. Model Numbers: DS2200X11B, IN2200X15B, BL2200X15B, BR2200X18B, CA2200X12B, DE2200X13B, GB2200X15B, ND2200X15B, ES2200X15B, FR2200X14B, FX2200X15B, EU2200X15B, IA2200X15B, IT2200X21B, KR2200X15B, TR2200X15B, EE2200X15B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.

    Lot or code
    Model Number (UDI-DI Number): DS2200X11B (606959052017); IN2200X15B (606959056497); BL2200X15B (606959058668); BR2200X18B (606959058583); CA2200X12B (606959052321); DE2200X13B (606959058651); GB2200X15B (606959058675); ND2200X15B (606959058620); ES2200X15B (606959058644); FR2200X14B (606959058613); FX2200X15B (606959061033); EU2200X15B (606959058682); IA2200X15B (606959058606); IT2200X21B (606959058637); KR2200X15B (606959058590); TR2200X15B (606959058798); EE2200X15B (606959058705). All Serial Numbers.
    FDA recall number
    Z-0293-2025
    Quantity
    34,146 units
  • Aeris EVO. Model Number: VT2110X24B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.

    Lot or code
    Model Number: VT2110X24B, UDI-DI Number: 606959062962. All Serial Numbers.
    FDA recall number
    Z-0294-2025
    Quantity
    804 units
  • Garbin Evo. Model Number: LD2110X23B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.

    Lot or code
    Model No. LD2110X23B; UDI-DI: 606959057432; All Serial Numbers.
    FDA recall number
    Z-0295-2025
    Quantity
    1,322 units
  • LifeVent EVO2. Model Number: SP2100X26B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.

    Lot or code
    Model No. SP2100X26B; UDI-DI: 606959057456; All Serial Numbers.
    FDA recall number
    Z-0296-2025
    Quantity
    1,244 units

Distribution (FDA): Worldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, LY, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, TZ, UA, UY, VN, YT, ZA, ZW.

Source

Issued by
U.S. Food and Drug Administration
Firm
Philips Respironics, Inc.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Philips Respironics, Inc., Month: November 2024, Week 2024-W47. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 95486 on November 20, 2024.

Other notices of Philips Respironics, Inc.