What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).
- Lot or code
- REF/UDI-DI: 1000/05425025750405, 1000-D/05425025750528 All Serial Numbers less than 500,000
- FDA recall number
- Z-1072-2025
- Quantity
- 81,606
SenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation. Used for Vagus Nerve Stimulation (VNS).
- Lot or code
- UDI-DI: 05425025750405. All Serial Numbers less than 500,000.
- FDA recall number
- Z-1073-2025
- Quantity
- 48
Distribution (FDA): US: AK, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PN, SC, SD, TN, TX, VA, VT, WA, WI, WV, UT. OUS: AE, AR, AT, BE, BG, BR, CA, CL, CO, CR, CZ, DE, DK, GR, HR, IL, IN, IT, JO, KW, LB, MK, NL, OM, PA, PE, PL, QA, RO, RS, SA, SE, SG, SI, TR, TT, UK, ZA.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- LivaNova USA, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.