Spacelabs Healthcare, Inc. recalls
6 notices naming Spacelabs Healthcare, Inc. as the recalling firm, since January 15, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Spacelabs Healthcare, Inc.: Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102…
Hazard (FDA): Due to off-the-shelf display port component that has audio/video failures
FDAMedical devices
Spacelabs Healthcare, Inc.: Brand Name: Spacelabs Ultraview SL Command Module Product Name: SL Command Module…
Hazard (FDA): Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).
FDAMedical devices
Spacelabs Healthcare, Inc.: Multi-parameter Command Module
Hazard (FDA): Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO…
FDAMedical devices
Spacelabs Healthcare, Inc.: Brand Name: Branding is not applied to this product. Product Name: DVI display cables… and 2 more
Hazard (FDA): Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of…
FDAMedical devices
Spacelabs Healthcare, Inc.: Xhibit Telemetry Receiver
Hazard (FDA): Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If…
FDAMedical devices
Spacelabs Healthcare, Inc.: Xhibit Telemetry Receiver with Software
Hazard (FDA): Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If…
FDAMedical devices