Shimadzu Corp. Analytical & Measuring Instruments Div.: Autosampler
Hazard (FDA): There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and…
FDAMedical devices
Shimadzu Corp. Analytical & Measuring Instruments Div.: Autosampler
Hazard (FDA): There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and…