Radiometer Medical ApS recalls
2 notices naming Radiometer Medical ApS as the recalling firm, since April 30, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Radiometer Medical ApS: The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists
Hazard (FDA): In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil, 2)…
FDAMedical devices
Radiometer Medical ApS: Radiometer ABL90 Series - Radiometer ABL90 FLEX Model Number 393-090 Radiometer ABL90…
Hazard (FDA): ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software…
FDAMedical devices