Ortho-Clinical Diagnostics, Inc. recalls
9 notices naming Ortho-Clinical Diagnostics, Inc. as the recalling firm, since December 4, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Ortho-Clinical Diagnostics, Inc.: Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains 5… and 1 more
Hazard (FDA): VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J…
FDAMedical devices
Ortho-Clinical Diagnostics, Inc.: VITROS Chemistry Products Ca Slides
Hazard (FDA): Generations 67 and above of VITROS Ca Slides are experiencing an increased occurrence of Condition Code TH4-63J during calibration on VITROS XT 3400 and VITROS XT 7600 Systems, causing calibration failures that delay…
FDAMedical devices
Ortho-Clinical Diagnostics, Inc.: Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products… and 3 more
Hazard (FDA): Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution…
FDAMedical devices
Ortho-Clinical Diagnostics, Inc.: VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP…
Hazard (FDA): Lots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than expected quality control (QC) results. If the customer is unable to get passing quality control results, then the customer…
FDAMedical devices
Ortho-Clinical Diagnostics, Inc.: Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: and 2 more
Hazard (FDA): A software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date…
FDAMedical devices
Ortho-Clinical Diagnostics, Inc.: Brand Name: Solana¿ Bordetella Complete Assay Product Name: Solana Bordetella kit…
Hazard (FDA): It was determined that there could be a false positive rate of up to 16.7% for the affected lots associated with the Bordetella parapertussis (BPP) results (only) of the assay.
FDAMedical devices
Ortho-Clinical Diagnostics, Inc.: VITROS Chemistry Products CRP Slides. One sales unit contains 5 cartridges
Hazard (FDA): Product was incorrectly assigned GEN 88 instead of GEN 83 on product carton, foil wrapper, and cartridge (cart) label.
FDAMedical devices
Ortho-Clinical Diagnostics, INc.: Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number… and 2 more
Hazard (FDA): The VITROS 5600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired…
FDAMedical devices
Ortho-Clinical Diagnostics, Inc.: VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures…
Hazard (FDA): May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L range) while the samples having higher than 90 mg/L. User no way to…
FDAMedical devices