Noah Medical recalls
3 notices naming Noah Medical as the recalling firm, since December 18, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Noah Medical: Galaxy System
Hazard (FDA): Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of…
FDAMedical devices
Noah Medical: Galaxy System
Hazard (FDA): Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is…
FDAMedical devices
Noah Medical: Galaxy bronchoscope
Hazard (FDA): Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive…
FDAMedical devices