Mint Medical GmbH recalls
2 notices naming Mint Medical GmbH as the recalling firm, since December 25, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Mint Medical GmbH: Mint Lesion
Hazard (FDA): If the connection between a mint Lesion workstation and the mint Lesion server is interrupted while performing a read, in rare cases some information may be lost or incorrectly linked to other patients.
FDAMedical devices
Mint Medical GmbH: mint Lesion and 1 more
Hazard (FDA): Some software versions have a malfunction where they may show incorrect orientation labels for a specific subset of DICOM images.
FDAMedical devices