Medical Action Industries, Inc. 306 recalls
6 notices naming Medical Action Industries, Inc. 306 as the recalling firm, since May 13, 2026. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Medical Action Industries, Inc. 306: Generic Name: Kit
Hazard (FDA): An internal data-entry error caused kit lot 333606 to be labeled with an expiration date of March 31, 2027, instead of February 28, 2027.
FDAMedical devices
Medical Action Industries, Inc. 306: Blood Culture Collection Venipuncture Kit 20/Cs
Hazard (FDA): Blood culture convenience kits contain a component subject to a drug recall that due to an an open or incomplete seal on the packaging of the applicator may compromise the component's sterile barrier and if used may…
FDAMedical devices
Sterility
Medical Action Industries, Inc. 306: Kit: Epidural/Pain TQC 20/Cs and 1 more
Hazard (FDA): Kits containing Sodium Chloride ampules subject to another recall are recalled due to potentially compromised sterility.
FDAMedical devices
Medical Action Industries, Inc. 306: Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence…
Hazard (FDA): Convenience kits contain needle free access device that has an improper seal which may result in doming or tearing of the silicone seal.
FDAMedical devices
Medical Action Industries, Inc. 306: Pack Angiography NCMC 3/Cs BANNER
Hazard (FDA): Cardiovascular procedure kits contain manifolds, the subject of another recall, that may have particulate within the fluid path, which has the potential to be introduced into blood circulation and become lodged within…
FDAMedical devices
Medical Action Industries, Inc. 306: Medical Action Industries Inc. Pack Cath BHH
Hazard (FDA): Convenience kit contains a syringe in which the rotating adapter may unwind during use, resulting in a loose connection and/or full disconnection between the syringe and manifold"