Karl Storz Endoscopy recalls
6 notices naming Karl Storz Endoscopy as the recalling firm, since November 20, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Karl Storz Endoscopy: HOPKINS Telescope 30¿
Hazard (FDA): Due to incorrect inside packaging label for eyepiece.
FDAMedical devices
Sterility
Karl Storz Endoscopy: Flexible Hystero-Fiberscope Kits REF: 11264BBU1
Hazard (FDA): Flexible hysteroscopes indicated for use during diagnostic and operative hysteroscopic procedures may contain an expired single use endoscopic seal for gynecology use, which could impact the assurance of sterility.
FDAMedical devices
Karl Storz Endoscopy: KARL STORZ - ENDOSKOPE and 13 more
Hazard (FDA): Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.
FDAMedical devices
Sterility
Karl Storz Endoscopy: Karl Storz - Endoskope - REF 723400 - Optical Biopsy and Grasping Forceps
Hazard (FDA): Due to failure of manual cleaning validation, biopsy and grasping forceps like sterility assurance.
FDAMedical devices
Sterility
Karl Storz Endoscopy: 4.5 x 180 mm REF 28208BKS Aggressive Cutter
Hazard (FDA): Due to holes detected in the inner barrier system that may compromise the sterility of the affected products.
FDAMedical devices
Karl Storz Endoscopy: Irrigation Tubing for KARL STORZ Endomat Select UP210 REF 031523-10 Tubing Set
Hazard (FDA): Due to improper labeling of products. An intended use on the label has not been reviewed and approved by FDA for distribution in the U.S.
FDAMedical devices