Hamilton Medical AG recalls
4 notices naming Hamilton Medical AG as the recalling firm, since August 6, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Hamilton Medical AG: IntelliCuff
Hazard (FDA): Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.
FDAMedical devices
Hamilton Medical AG: Breathing circuit set
Hazard (FDA): Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may…
FDAMedical devices
Hamilton Medical AG: Hamilton C6 REF 160021
Hazard (FDA): A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device…
FDAMedical devices
Hamilton Medical AG: Breathing circuit set
Hazard (FDA): Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete…
FDAMedical devices