Contamination
Greiner Bio-One North America, Inc.: Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number…
Hazard (FDA): Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the…
FDAMedical devices
Greiner Bio-One North America, Inc.: Brand Name: Vacuette Product Name: 8 ml CAT Serum Separator Clot Activator Model/Catalog…
Hazard (FDA): Some of the Vacuette¿ CAT Serum Separator Clot Activator tubes were manufactured without gel separator, which could yield erroneous assay results and require the patient to return to the lab for recollection or repeat…
FDAMedical devices
Greiner Bio-One North America, Inc.: Greiner BIO-ONE VACUETTE 2 ml 9NC Coagulation sodium citrate 3.2%
Hazard (FDA): Blood collection tubes were manufactured with the label applied too high on the tube, which shifted the printed fill level mark upward, which could result in unnecessary sample rejection and repeat collections.