Getinge Disinfection Ab recalls
2 notices naming Getinge Disinfection Ab as the recalling firm, since May 14, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Getinge Disinfection Ab: Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
Hazard (FDA): Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
FDAMedical devices
Getinge Disinfection Ab: Washer Disinfector Aquadis 56 and 1 more
Hazard (FDA): The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the…
FDAMedical devices