GE Medical Systems China Co., Ltd. recalls
4 notices naming GE Medical Systems China Co., Ltd. as the recalling firm, since May 7, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
GE Medical Systems China Co., Ltd.: MAC 5 A4 Resting ECG Analysis System 855001-001
Hazard (FDA): Software in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical…
FDAMedical devices
GE Medical Systems China Co., Ltd.: GE Healthcare Carestation anesthesia system and 2 more
Hazard (FDA): Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia…
FDAMedical devices
GE Medical Systems China Co., Ltd.: GE Healthcare LOGIQ F R3 Ultrasound System and 1 more
Hazard (FDA): GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz…
FDAMedical devices
GE Medical Systems China Co., Ltd.: GE Healthcare CARESTATION 620 A1 and 14 more
Hazard (FDA): Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in…
FDAMedical devices