DeRoyal Industries Inc recalls
4 notices naming DeRoyal Industries Inc as the recalling firm, since February 26, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
DeRoyal Industries Inc: REF: 81-030415 Myocardial Probe Non-Pyrogenic
Hazard (FDA): Myocardial temperature probes contain an incorrect needle. Specifically, an 18mm needle was included instead of a 15mm needle.
FDAMedical devices
Sterility
DeRoyal Industries Inc: DeRoyal Industries Inc. Ref 46-406
Hazard (FDA): Sterile wound dressing, lacks sterility assurance
FDAMedical devices
DeRoyal Industries Inc: DeRoyal REF 71-C5100NS Tubing
Hazard (FDA): Due to the inner diameter of the suction tubing connector being smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a…
FDAMedical devices
DeRoyal Industries Inc: DeRoyal AMG Suction Connector Tubing REF: 71-2083
Hazard (FDA): The tubing may not ensure that the yankauer stays intact, and secure to the tubing / yankauer connection for a tight seal. This could result in inadequate suction and cause delay in patient care.
FDAMedical devices