Daavlin Distributing Company: Photera 400 and 2 more
Hazard (FDA): Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
FDAMedical devices
Daavlin Distributing Company: Phothera 200
Hazard (FDA): Some of the Calibration/Output Certificates had the outputs reversed.