Covidien: Covidien Nellcor Bedside SpO2 Patient Monitoring System: REF: PM100N
Hazard (FDA): Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient…
FDAMedical devices
Covidien: Newport HT70 Ventilator
Hazard (FDA): Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation…
FDAMedical devices
Covidien: Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula
Hazard (FDA): Tracheostomy tube with cuff reusable inner cannula has a flange that may disconnect from the outer cannula of the device, which could cause respiratory failure, aspiration, unspecified tissue injury, aspiration…
FDAMedical devices