Covidien, LLC recalls
5 notices naming Covidien, LLC as the recalling firm, since March 18, 2026. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Covidien, LLC: EndoFLIP System that includes: ENDOFLIP 1 SYSTEM EF-100 05391530810050 10884521809307…
Hazard (FDA): The system's 24-hour usage limit results in catheters incorrectly flagged as expired, making them unavailable for clinical procedures.
FDAMedical devices
Covidien LLC: Covidien Shiley Disposable Decannulation Plug REF DDCP UDI-DI code: 10884522004060…
Hazard (FDA): Due to product labeling incorrectly including the symbol indicating non-DEHP (Bis(2-ethylhexyl) phthalate as the product includes 30 to 35% DEHP.
FDAMedical devices
Covidien LLC: Covidien Mon-a-Therm Esophageal Stethoscope with Temperature Sensor 400TM REF: 90042 and 2 more
Hazard (FDA): Temperature probe devices lack FDA clearance.
FDAMedical devices
Covidien LLC: COVIDIEN Shiley Tracheostomy Tube Cuffed with Inner Cannula REF 6FEN UDI-DI code…
Hazard (FDA): Due to customer complaint regarding incorrect display box labeling.
FDAMedical devices
Covidien LLC: HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: Product…
Hazard (FDA): Due to complaints and investigations stating that alarms from the primary patient bedside monitor was not alarming/being transmitted and received on the remote patient monitoring system.
FDAMedical devices