Cardinal Health 200, LLC recalls
17 notices naming Cardinal Health 200, LLC as the recalling firm, since November 20, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Cardinal Health 200, LLC: Cardinal Health Presource PBDS kits and 10 more
Hazard (FDA): Kits contain infusion sets which were recalled for performance issues.
FDAMedical devices
Cardinal Health 200, LLC: Presource Packs containing Medtronic Microstream Components: Intubation Kit Catalog…
Hazard (FDA): Kits contain Microstream components. Medtronic issued an update to the Instructions for Use of these components due reports of difficulty in disconnecting the adapter from the patient's endotracheal tube to perform a…
FDAMedical devices
Cardinal Health 200, LLC: Presource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT and 3 more
Hazard (FDA): The kits contain tube sets with handpieces that may leak during use. The impacted devices can have the following failures: fluid leakage from the hand piece that can pool on the floor. The impact can range from user…
FDAMedical devices
Contaminated
Cardinal Health 200, LLC: Presource Convenience Kits: 1) KIT and 1 more
Hazard (FDA): The kits contain applicators that may exhibit an open or incomplete seal on the packaging of the applicator. Use of a potentially contaminated applicator could result in no patient harm or may lead to localized skin or…
FDAMedical devices
Contaminated
Cardinal Health 200, LLC: Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays… and 11 more
Hazard (FDA): The Urinary Catheter Trays contain CliniClean Pericare Wipes which may potentially be contaminated with a lubricant used during the manufacturing process. If a contaminated wipe is used on a patient, localized skin…
FDAMedical devices
Cardinal Health 200, LLC: Cardinal Health Presource Kits: 1) FAT GRAFT ADD KIT and 7 more
Hazard (FDA): Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
FDAMedical devices
Cardinal Health 200, LLC: Cardinal Health Monoject Tuberculin Syringe
Hazard (FDA): The outer packaging and blister pack labeling identify the product as 1mL Luer Lock Tuberculin Syringe (Product Code 1180100777), however the syringes contained within the packages are U-100 Insulin syringes.
FDAMedical devices
Sterility
Cardinal Health 200, LLC: Webcol Alcohol Prep pad
Hazard (FDA): Non-Sterility: microbial contamination identified as Paenibacillus phoenicis
FDAMedicines
Cardinal Health 200, LLC: Chest Drainage Units and Accessories: Product Code Description 1. 8888571562 Cardinal…
Hazard (FDA): The firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants, the CDU may not demonstrate…
FDAMedical devices
Mislabeled
Cardinal Health 200, LLC: ChemoPlus gowns: Product Code Product Description CT5500T ChemoPlus Full Coverage Sleeve
Hazard (FDA): Affected gowns have the incorrect expiration date on the product packaging. Product has a shelf-life of three years which is mislabeled on the product packaging as five (5) years.
FDAMedical devices
Cardinal Health 200, LLC: Salem Sump PVC Tubes: Product Code/Product Name: 8888264911E ENFit" SALEM SUMP" PVC TUBE… and 5 more
Hazard (FDA): Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
FDAMedical devices
Cardinal Health 200, LLC: SCD Comfort Knee Length Small and 2 more
Hazard (FDA): Incorrect software logic may cause frequent E1 errors (system high pressure). Issue may cause delay in treatment or therapy.
FDAMedical devices
Sterility
Cardinal Health 200, LLC: Catalog No. 8888160119 and 5 more
Hazard (FDA): Packaging defect may compromise sterility of the product. Use of a non-sterile product on the patient could lead to infection in neonates.
FDAMedical devices
Contamination
Cardinal Health 200, LLC: Cardinal Health Presource Kits: 1) KIT and 11 more
Hazard (FDA): Potential for endotoxin contamination in non-sterile surgical strips and patties in Presource Kits.
FDAMedical devices
Contamination
Cardinal Health 200, LLC: Cardinal Health Presource Kits: 1) KIT and 9 more
Hazard (FDA): Potential endotoxin contamination in non-sterile surgical strips and patties in Presource kits.
FDAMedical devices
Cardinal Health 200, LLC: Convertor Isolation Gown and 1 more
Hazard (FDA): Isolation Gowns were manufactured with raw materials that may fail the hydrostatic pressure specification and therefore would not comply with AAMI (The Association for the Advancement of Medical Instrumentation) Level…
FDAMedical devices
Cardinal Health 200, LLC: Monoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack
Hazard (FDA): Certain lots of listed lots of Cardinal Health Monoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack have demonstrated incompatibility with needleless Intravenous (IV) connectors. Cardinal Health…
FDAMedical devices