Beckman Coulter Mishima K.K. recalls
3 notices naming Beckman Coulter Mishima K.K. as the recalling firm, since June 11, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Beckman Coulter Mishima K.K.: Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480…
Hazard (FDA): Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC…
FDAMedical devices
Beckman Coulter Mishima K.K.: DxC 700 AU and 1 more
Hazard (FDA): A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases…
FDAMedical devices
Beckman Coulter Mishima K.K.: Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate… and 1 more
Hazard (FDA): A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing…
FDAMedical devices