Augustine Temperature Management, LLC: Hot Dog PATIENT WARMING and 2 more
Hazard (FDA): There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of…
FDAMedical devices
Augustine Temperature Management, LLC: Augustine Surgical Inc. HotDog Return Electrode accessory connector cables and 1 more
Hazard (FDA): Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .
FDAMedical devices
Augustine Temperature Management, LLC: Augustine Surgical Inc. HOT DOG PATIENT WARMING and 4 more
Hazard (FDA): HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
FDAMedical devices
Augustine Temperature Management, LLC: Augustine Surgical Inc. HOT DOG PATIENT WARMING and 2 more
Hazard (FDA): Instructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 pertaining to Power-on self-test error code "EA POST".