Accuray Incorporated: Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer
Hazard (FDA): In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room…
FDAMedical devices
Accuray Incorporated: Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer
Hazard (FDA): In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room…
FDAMedical devices
Accuray Incorporated: CyberKnife Treatment Delivery System
Hazard (FDA): Snap rings may become partially or completed detached from the shafts within the internal mechanism. Uncontrolled rotation may occur on the left and right roll axes when the rings are removed or damaged.